8. SDU5600 Command List
The SDU5600 can be operated via PC using the RS232 port.
8-1 Communication parameter and connecting lead
The SDU5600 and a PC should be connected using a straight cable between
the [PC I/F] socket and the serial port of the PC. The SDU5600 port is MALE,
so the PC lead must have a female connector or a gender changer will have to
be employed.
Communications parameters of the SDU5600
Baud rate 9600bps
Data length 8 bit
Parity None
Stop bit 2 bit
Flow control RTS/CTS
8-2 Delimiter
A brief description of delimeter and response format is given here.
Delimiter
Each command from the PC to SDU5600 is defined by:
<CR> 0x0d in 16-bit form
or
<CR><LF> 0x0d,0x0a in hexadecimal format
Note: <LF> will be ignored
Response when the command has been correct:
<SP><CR><LF> 0x20,0x0d,0x0a in hexadecimal format
Response when the command has been incorrect:
?<CR><LF> 0x3f,0x0d,0x0a in hexadecimal format
Response to the read command:
Following the output of the parameter, the correct response should read
<SP><CR><LF> 0x20,0x0d,0x0a in hexadecimal format
Section 8, 8-1, 8-2
Page view 38